Early Radiographic and Clinical Outcomes of an Additive-Manufactured Acetabular Component
Logan Radtke1, Jeffrey J Frandsen1, Alex J Lancaster1
1Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT, USA.
Background:
Additive manufacturing has recently gained popularity and is widely adopted in the orthopaedic industry. However, there is a paucity of literature on the radiographic and clinical outcomes of these relatively novel components. The aim of this study was to assess the 2-year clinical and radiographic outcomes of a specific additive-manufactured acetabular component in primary total hip arthroplasty.
Methods:
We performed a retrospective review of 60 patients who underwent primary total hip arthroplasty with the use of the Stryker's TRIDENT II acetabular component. Evaluation of radiographs was performed at 6 weeks, 1 year, and 2 years postoperatively. Radiographs were evaluated for radiolucencies in Charnley and DeLee zones, signs of biologic fixation, and acetabular inclination and anteversion measurements. Patient-reported outcomes and complications were also obtained.
Results:
There were no cases of component loosening or changes in component position during follow-up, with an average follow-up time of 1.7 years. A radiolucent line was identified in one patient in zone 1 at 6 weeks; this was absent at 1 year. Radiographic signs of cup biologic fixation were present in 85% of cases by final follow-up. The average inclination was 45.1 (SD = 4.0), and the average anteversion was 26.9 (SD = 5.2). Patient-Reported Outcomes Measurement Information System scores significantly increased at the final follow-up, and there were no complications in this cohort.
Conclusions:
This study demonstrated excellent radiographic and clinical outcomes with this novel additive-manufactured acetabular component at early follow-up. Although longer-term follow-up is warranted, this additively manufactured highly porous titanium acetabular component demonstrated excellent biologic fixation and reliable fixation at mid-term follow-up.