Related Experiment Video
Updated: Jun 29, 2025

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Cost analysis of different medical oxygen sources for a healthcare facility in India
Varun Manhas1, Mahipal Rawat2, Yogesh S Kaurav3
1One Health Trust, Bengaluru, Karnataka, India.
Background And Aims:
Multiple sources of medical oxygen, namely liquid medical oxygen (LMO) tanks, pressure swing adsorption (PSA) plants, concentrators, and gaseous cylinders, are available at different healthcare facilities. These sources of oxygen have varying installation and operational costs. In low-resource settings, it is imperative to utilise these assets optimally. This study investigated the operational costs of multiple oxygen sources available at a healthcare facility.
Methods:
A Microsoft (MS) Excel-based model was developed to analyse and compare the oxygen manufacturing costs (in ₹/m3) using PSA plants and procurement costs (in ₹/m3) of LMO and third-party vendor-refilled cylinders.
Results:
The oxygen manufacturing costs for PSA plants of different capacities and running times on electricity and diesel generators (DGs) as a power source were calculated. This study highlights the cost-benefit of using PSA plants over LMO and third-party vendor-refilled cylinders as a source of oxygen. PSA plants are most economical when they are of higher capacity and used to their maximum capacity on electricity as the power source. On the contrary, they are most expensive when used on a DG set as a power source. Furthermore, this study provides evidence of PSA plants being more cost-effective for refilling cylinders using a booster compressor unit when compared to third-party vendor-cylinder refilling.
Conclusion:
Given their cost-effectiveness and low third-party dependence, they should be utilised to their maximum capacity as medical oxygen sources at healthcare facilities.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
07:38Design and Development of a Model to Study the Effect of Supplemental Oxygen on the Cystic Fibrosis Airway Microbiome
Published on: August 3, 2021
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...