Non-contrast MRI and post-mastectomy silicone breast implant rupture: preventing false positive diagnoses
Luca Mazzocconi1, Francesca De Lorenzi, Riccardo Carbonaro
1European Institute of Oncology, IRCCS, Via Ripamonti, Milan, Italy.
Background:
Breast implants are not lifelong, with implant rupture being the third leading cause of revisional surgery in augmented women. Noncontrast MRI is a reliable tool to assess implant integrity; however, false positive and false negative diagnoses have been reported due to an incorrect interpretation of MRI signs. This study aims to investigate the incidence of these misleading results, comparing MRI findings with intraoperative surgical observations and exploring signs of nonunivocal interpretation.
Materials And Methods:
Between March 2019 and October 2022, our hospital, a referral center for breast cancer care, conducted 139 breast MRI examinations to evaluate implant integrity. Surgical intervention was deemed necessary for patients diagnosed with suspected or confirmed implant rupture at MRI. Those patients who did not undergo any surgical procedure (63 cases) or had surgery at different institutes (11 cases) were excluded.
Results:
Among the 65 patients who underwent preoperative MRI and subsequent surgery at our institute, surgical findings confirmed the preoperative MRI diagnosis in 48 women. Notably, 17 women exhibited a discordance between MRI and surgical findings: three false negatives, 11 false positives and three possible ruptures not confirmed. Signs of nonunivocal or misleading interpretation were assessed on a patient-by-patient basis. The importance of obtaining detailed information about a patient's breast implant, including fill materials, number of lumens, manufacturer and shape, proved immensely beneficial for interpreting MRI signs accurately.
Conclusion:
Pre-MRI knowledge of implant details and a meticulous evaluation of non-univocal signs can aid radiologists in accurately assessing implant integrity, reducing the risk of unnecessary revisional surgeries, and potentially averting allegations of medical malpractice.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
13:44Detection of Architectural Distortion in Prior Mammograms via Analysis of Oriented Patterns
Published on: August 30, 2013
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...