Segmented RF shield design to minimize eddy currents for low-field Halbach MRI systems
Bart de Vos1, Rob Remis2, Andrew Webb1
1C.J. Gorter MRI center, Radiology, Leiden University Center, Leiden, The Netherlands; Terahertz Sensing, Microelectronics, Delft University of Technology, Delft, The Netherlands.
Abstract:
MRI systems have a thin conducting layer placed between the gradient and RF coils, this acts as a shield at the RF-frequency, minimizing noise coupled into the experiment, and decreasing the coupling between the RF and gradient coils. Ideally, this layer should be transparent to the gradient fields to reduce eddy currents. In this work the design of such a shield, specifically for low-field point-of-care Halbach based MRI devices, is discussed. A segmented double layer shield is designed and constructed based on eddy current simulations. Subsequently, the performance of the improved shield is compared to a reference shield by measuring the eddy current decay times as well as using noise measurements. A maximum reduction factor of 2.9 in the eddy current decay time is observed. The segmented shield couples in an equivalent amount of noise when compared to the unsegmented reference shield. Turbo spin echo images of a phantom and the brain of a healthy volunteer show improvements in terms of blurring using the segmented shield.
Related Concept Videos
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Magnetic Field Of A Current Loop
Magnetic Resonance Imaging
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
NMR Spectrometers: Resolution and Error Correction