Comparison of 3T and 7T magnetic resonance imaging for direct visualization of finger flexor pulley rupture: an
Thomas Bayer1,2, Lilly Bächter3, Christoph Lutter4,5
1Institue of Radiology, Universitätsklinikum & Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. thomas.bayer@klinikum-fuerth.de.
Objective:
To compare image quality and diagnostic performance of 3T and 7T magnetic resonance imaging (MRI) for direct depiction of finger flexor pulleys A2, A3 and A4 before and after artificial pulley rupture in an ex-vivo model using anatomic preparation as reference.
Materials And Methods:
30 fingers from 10 human cadavers were examined at 3T and 7T before and after being subjected to iatrogenic pulley rupture. MRI protocols were comparable in duration, both lasting less than 22 min. Two experienced radiologists evaluated the MRIs. Image quality was graded according to a 4-point Likert scale. Anatomic preparation was used as gold standard.
Results:
In comparison, 7T versus 3T had a sensitivity and specificity for the detection of A2, A3 and A4 pulley lesions with 100% vs. 95%, respectively 98% vs. 100%. In the assessment of A3 pulley lesions sensitivity of 7T was superior to 3T MRI (100% vs. 83%), whereas specificity was lower (95% vs. 100%). Image quality assessed before and after iatrogenic rupture was comparable with 2.74 for 7T and 2.61 for 3T. Visualization of the A3 finger flexor pulley before rupture creation was significantly better for 7 T (p < 0.001). Interobserver variability showed substantial agreement at 3T (κ = 0.80) and almost perfect agreement at 7T (κ = 0.90).
Conclusion:
MRI at 3T allows a comparable diagnostic performance to 7T for direct visualization and characterization of finger flexor pulleys before and after rupture, with superiority of 7T MRI in the visualization of the normal A3 pulley.
Insights
High-field 7 Tesla (T) and 3 T magnetic resonance imaging (MRI) show comparable diagnostic performance for finger flexor pulleys. 7T MRI offers superior visualization of the normal A3 pulley, with both modalities effective in detecting pulley ruptures.
Area of Science:
- Musculoskeletal Imaging
- Advanced MRI Techniques
- Anatomical Visualization
Background:
- Direct visualization of finger flexor pulleys (A2, A3, A4) is crucial for diagnosing injuries.
- Magnetic Resonance Imaging (MRI) at different field strengths (3T and 7T) offers potential for enhanced imaging.
- Ex-vivo models provide a controlled environment for evaluating imaging techniques.
Purpose of the Study:
- To compare the image quality and diagnostic performance of 3T and 7T MRI.
- To assess the depiction of finger flexor pulleys A2, A3, and A4.
- To evaluate imaging before and after induced pulley rupture using anatomic preparation as a reference.
Main Methods:
- 30 human cadaveric fingers were imaged at 3T and 7T MRI.
- Fingers underwent iatrogenic pulley rupture prior to a second imaging session.
- Two radiologists assessed image quality and diagnostic performance, with anatomic preparation as the gold standard.
Main Results:
- Sensitivity and specificity for pulley lesions were high for both 7T (100%/98%) and 3T (95%/100%).
- 7T MRI demonstrated superior sensitivity (100% vs. 83%) for A3 pulley lesions, though with lower specificity (95% vs. 100%).
- Image quality was comparable (7T: 2.74, 3T: 2.61); 7T showed significantly better visualization of the normal A3 pulley (p<0.001) with almost perfect interobserver agreement (κ=0.90 vs. 0.80 at 3T).
Conclusions:
- 3T MRI provides diagnostic performance comparable to 7T MRI for finger flexor pulleys.
- 7T MRI exhibits superiority in visualizing the normal A3 pulley.
- Both 3T and 7T MRI are effective for characterizing finger flexor pulleys before and after rupture.