Purpose:
Continuous shear wave elastography (C-SWE) can be expected to be applied to portable muscle elasticity diagnosis. To establish diagnostic technology, it will be necessary to improve measurement techniques and quantitative measurement accuracy.
Methods:
In this study, we investigated two screen scores: the quality index (Q-index), which determines whether the intensity of a power Doppler image is appropriate, and the shear wave propagation direction index (SWDI), which determines the uniformity of shear wave propagation.
Results:
First, we performed numerical simulations with white noise and found that the coefficient of variation of shear wave velocity estimation was less than 5% when the normalized Q-index was greater than 0.27. Furthermore, regarding the SWDI, we clarified the relationship between the standard deviation in shear wave propagation direction and the SWDI. Next, the relationship between the Q-index and coefficient of variation of estimated shear wave velocity was evaluated through experiments using a tissue-mimicking phantom. The results showed that there was a negative correlation between the Q-index and the coefficient of variation, and the fluctuation of the propagation velocity could be inferred from the Q-index. Finally, we showed the results of applying the screen scores to muscle relaxation monitoring and confirmed its usefulness in clinical applications.
Conclusion:
By applying the screen scores, we showed improved stability in speed estimation in C-SWE, and demonstrated the possibility of clinical applicability.