Pressure and stiffness of a new air-inflated compression wrap
Giovanni Mosti1, Antonios Gasparis2, Alisha Oropallo2
1Angiology Department, Clinica MD Barbantini, Lucca, Italy.
Objective:
To report pressure and stiffness, in healthy volunteers, of a new compression device with an air bladder inflated by a pump to regulate pressure.
Methods:
The device was applied to 60 legs of 30 volunteers and set to exert different pressures of 20-50 mmHg. The exerted pressure was measured in supine and standing positions and during simple physical exercises; static stiffness index, dynamic stiffness index, and walking pressure amplitudes were calculated.
Results:
The exerted pressure showed a good correlation with the expected pressure at each pressure range. The stiffness indices were >10 mmHg in the range of inelastic materials. The device was considered very easy to apply and use by the testing researchers.
Conclusions:
The device stiffness is in the same range as the inelastic bandages. Consequently, similar hemodynamic effectiveness could be expected but must be proved. Unlike inelastic bandages, this device was easy to apply and use.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Residual Stresses in Bending