Aims:
A three-dimensional electroanatomic mapping system-guided transseptal puncture (3D-TSP), without fluoroscopy or echocardiography, has been only minimally reported. Indications for 3D-TSP remain unclear. Against this background, this study aims to establish a precise technique and create a workflow for validating and selecting eligible patients for fluoroless 3D-TSP.
Methods And Results:
We developed a new methodology for 3D-TSP based on a unipolar electrogram derived from a transseptal needle tip (UEGM tip) in 102 patients (the derivation cohort) with intracardiac echocardiography (ICE) from March 2018 to February 2019. The apparent current of injury (COI) was recorded at the muscular limbus of the foramen ovalis (FO) on the UEGM tip (sinus rhythm: 2.57 ± 0.95 mV, atrial fibrillation: 1.92 ± 0.77 mV), which then disappeared or significantly reduced at the central FO. Changes in the COI, serving as a major criterion to establish a 3D-TSP workflow, proved to be the most valuable indicator for identifying the FO in 99% (101/102) of patients compared with three previous techniques (three minor criteria) of reduction in atrial unipolar or bipolar potential and FO protrusion. A total of 99.9% (1042/1043) patients in the validation cohort underwent successful 3D-TSP through the workflow from March 2019 to July 2023. Intracardiac echocardiography guidance was required for 6.6% (69/1042) of patients. All four criteria were met in 740 patients, resulting in a 100% pure fluoroless 3D-TSP success rate.
Conclusion:
In most patients, fluoroless 3D-TSP was successfully achieved using changes in the COI on the UEGM tip. Patients who met all four criteria were considered suitable for 3D-TSP, while those who met none required ICE guidance.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
08:31A Simplified Stepwise Approach to Echo Guidance during Percutaneous Mitral Valve Repair
Published on: October 16, 2021