Related Experiment Video
Updated: Jun 28, 2025

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
A novel negative pressure isolation device reduces aerosol exposure: A randomized controlled trial
John S Shin1, Muchuan Niu2, Haoxuan Chen2
1Department of Anesthesiology and Perioperative Medicine, University of California Los Angeles (UCLA), Los Angeles, CA, USA.
The Suction-Assisted Local Aerosol Containment Chamber (SLACC) effectively reduces healthcare worker exposure to airborne particles during intubation. This airway management device proved safe and did not hinder the intubation process in a clinical trial.
Area of Science:
- Anesthesiology
- Infectious Disease Control
- Medical Device Development
Background:
- The COVID-19 pandemic increased the need for intubation barriers to protect healthcare workers.
- The Suction-Assisted Local Aerosol Containment Chamber (SLACC) was developed as a novel intubation device.
- Evaluating SLACC's safety, ease of use, and aerosol containment efficacy was crucial.
Purpose of the Study:
- To assess the safety and ease of airway management using the SLACC device.
- To quantify the aerosol containment efficacy of SLACC in reducing healthcare worker exposure.
- To compare particle exposure during airway management with and without SLACC.
Main Methods:
- A randomized clinical trial involving adult patients undergoing elective surgery with general endotracheal anesthesia.
- Patients were randomized to airway management with or without the SLACC device.
- Nebulized saline was used to simulate infectious aerosol particles during anesthesia induction.
Main Results:
- 79 patients were enrolled; all intubations were successful with no difference in intubation time between groups.
- Healthcare workers experienced significantly lower particle number concentrations when using SLACC.
- SLACC reduced particle counts outside the device by 97% compared to counts inside.
Conclusions:
- The SLACC device is safe and does not impede airway management procedures.
- SLACC significantly reduces healthcare worker exposure to aerosolized particles during intubation.
- SLACC demonstrates high efficacy in containing infectious aerosols during airway procedures.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Suctioning the Nasopharyngeal Airway
Equipment Required