Related Experiment Video
Updated: Jun 28, 2025

Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Interchanging Reusable and Disposable Nebulizers Used with Home-Based Compressors May Result in Inconsistent Dosing:
Judy Schloss1, Dominic P Coppolo1, Jason A Suggett2
1Monaghan Medical Corporation, Plattsburgh, NY, 12901, USA.
Introduction:
Reusable nebulizer-compressor combinations deliver inhaled medications for patients with chronic lung diseases. On hospital discharge, the patient may take home the disposable nebulizer that was packaged and combine it with their home compressor. Though this practice may reduce waste, it can increase variability in medication delivery. Our study compared several reusable and disposable nebulizers packaged with compressor kits used in the US. We included a common disposable hospital nebulizer that may not be supplied with popular home kits but may be brought home after a hospitalization or emergency department visit. We focused on fine droplet mass < 4.7 μm aerodynamic diameter (FDM<4.7 μm), associated with medication delivery to the airways of the lungs.
Methods:
We evaluated the following nebulizer-compressor combinations (n = 5 replicates): 1. OMBRA® Table Top Compressor with MC 300® reusable and Airlife™ MistyMax™ 10® disposable nebulizer, 2. Sami-the-Seal® compressor with SideStream® reusable and disposable nebulizers and Airlife™ MistyMax 10™ disposable nebulizer, 3. VIOS® compressor with LC Sprint® reusable, and VixOne® and Airlife™ MistyMax™ disposable nebulizers, 4. Innospire® Elegance® compressor with SideStream® reusable and disposable nebulizers and Airlife™ MistyMax 10™ disposable nebulizer, 5. Willis-the-Whale® compressor with SideStream® reusable and disposable nebulizers and Airlife™ MistyMax 10™ disposable nebulizer, 6. Pari PRONEB® Max compressor with LC Sprint® reusable and Airlife™ MistyMax 10™ disposable nebulizer. We placed a 3-ml albuterol solution (0.833 mg/ml) in each nebulizer. A bacterial/viral filter was attached to the nebulizer mouthpiece to capture emitted medication, with the filter exit coupled to a simulator of a tidal breathing adult (rate = 10 cycles/min; Vt = 600 ml; I/E ratio = 1:2). The filter was replaced at 1-min intervals until onset of sputter. Droplet size distributions (n = 5 replicates/system) were determined in parallel by laser diffractometry.
Results:
Cumulative FDM<4.7 μm varied from 381 ± 33 μg for the best performing combination (Proneb/LC-Sprint) to 150 ± 21 μg for the system with the lowest output (VIOS®/MistyMax 10™).
Conclusions:
Substituting one nebulizer for another can result in large differences in medication delivery to the lungs.
Related Concept Videos
Inhaled Medications
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...