Tear meniscus height agreement and reproducibility between two corneal topographers and spectral-domain optical
Ivo Soares1, Eva Ramalho1, Francisco Miguel Brardo1
1Department of Physics, University of Beira Interior, Covilhã, Portugal.
Clinical Relevance:
Tear meniscus height (TMH) is an important clinical marker in dry eye diagnosis and management.
Purpose:
To evaluate the reproducibility and agreement of TMH measurements in non-clinical participants using the Oculus Keratograph 5 M, Medmont Meridia, and Spectral-domain optical coherence tomography (Spectralis SD-OCT).
Methods:
Fifty-six participants (mean 43.8 ± 22.4 years) were recruited for this cross-sectional study. Image acquisitions were performed on the three devices, sequentially and randomized. The repeatability and reproducibility of inter-observer and inter-device analysis were performed. Repeated measures ANOVA and Bland-Altman Plots were used to evaluate the agreement between devices.
Results:
The mean TMH with the Oculus Keratograph 5 M, Medmont Meridia and Spectralis SD-OCT were 0.29 ± 0.16 mm, 0.24 ± 0.09 mm and 0.27 ± 0.16 mm, respectively. There were no significant inter-observer differences (paired t-tests, p < 0.001). All the devices exhibited good inter-observer reliability (ICC ≥ 0.877), and good repeatability (CV ≤ 16.53%). Inter-device reliability is moderate (ICC = 0.621, p < 0.001). Repeated measures ANOVA revealed that TMH measurements given by the Spectralis SD-OCT are not significantly different from the Oculus Keratograph 5 M (p = 0.19) and the Medmont Meridia (p = 0.38). TMH measurements from Oculus Keratograph 5 M were significantly higher than those from Medmont Meridia (p = 0.02). Correlations between the mean TMH and the difference in the TMH measurements were positive for Oculus Keratograph 5 M and Medmont Meridia (r2 = 0.62, p < 0.001), negative for Medmont Meridia and Spectralis SD-OCT (r2 = -0.59, p < 0.001), and not significant for Oculus Keratograph 5 M and Spectralis SD-OCT (r2 = 0.05, p = 0.74). A strong correlation was found for TMH measured with all devices (r2 = 0.55 to 0.81, p < 0.001).
Conclusions:
The Oculus Keratograph 5 M, Medmont Meridia, and Spectralis SD-OCT provide reliable and reproducible inter-observer TMH measurements. Inter-device reliability is moderate, with a close correlation between Spectralis SD-OCT and the Oculus Keratograph 5 M. Oculus Keratograph 5 M and Medmont Meridia are repeatable devices appropriate for the measurement of TMH, but they are not interchangeable in clinical practice.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...