Binocular Dynamic Visual Acuity in Eyeglass-Corrected Myopic Patients
Published on: March 29, 2022
Evaluation of objective and subjective binocular ocular refraction with looking in type
Megumi Fukushima1, Masakazu Hirota2,3, Takafumi Yukimori4
1Division of Orthoptics, Graduate School of Medical Care and Technology, Teikyo University, Itabashi, Tokyo, Japan.
Background:
This study aimed to compare the results of the Chronos binocular/monocular refraction system, that measures objective and subjective ocular refraction in one unit, to objective findings obtained from a conventional autorefractometer and a conventional subjective ocular refraction using a trial-frame in real space.
Methods:
Twenty-eight healthy volunteers (21.2 ± 1.5 years old) were included in this study. Objective ocular refraction was measured using two tests: the Chronos binocular/monocular refraction system under binocular conditions and a conventional autorefractometer under monocular conditions. Subjective ocular refraction was measured using three tests: Chronos binocular/monocular refraction system under binocular, monocular conditions, and trial-frame in the real space under monocular conditions. The measurement distance was set to 5.0 m for each test. All ocular refractions were converted into spherical equivalents (SEs).
Results:
The objective SE was significantly more negative with Chronos binocular/monocular refraction system under binocular condition (- 4.08 ± 2.76 D) than with the conventional autorefractometer under monocular condition (- 3.85 ± 2.66 D) (P = 0.002). Although, the subjective SE was significantly more negative with Chronos binocular/monocular refraction system under binocular condition (- 3.55 ± 2.67 D) than with the trial-frame in the real space under monocular condition (- 3.33 ± 2.75 D) (P = 0.002), Chronos binocular/monocular refraction system under monocular condition (- 3.17 ± 2.57 D) was not significantly different from that in trial-frame in real space under monocular condition (P = 0.33).
Conclusion:
These findings suggest that the Chronos binocular/monocular refraction system, which can complete both objective and subjective ocular refraction tests in a single unit, is suitable for screening ocular refraction, although it produces slightly more myopic results. Furthermore, subjective ocular refraction testing accuracy in Chronos binocular/monocular refraction system can be equivalent to trial-frame in real-space testing by switching from binocular to monocular condition.