A New Biomechanical Deformation Response Parameter: Change in Central Corneal Thickness During Air Puff Induced
Monica D Okon1, Yanhui Ma2, B Audrey Nguyen-Rudy1
1Department of Biomedical Engineering, The Ohio State University, Columbus, OH, USA.
Purpose:
To investigate the percent change in central corneal thickness (%ΔCCT) during air-puff-induced deformation as an indicator of corneal biomechanical response.
Methods:
Forty ex vivo human eyes from forty donors were imaged using the CorVis ST at experimentally controlled intraocular pressure (IOP) of 10, 20, 30, and 40 mmHg, followed by uniaxial strip testing to calculate tensile modulus. The CorVis ST research software tracked the anterior and posterior cornea edges and determined the dynamic corneal response (DCR) parameters. Eyes were excluded if image quality or posterior tracking issues were present. Custom algorithms were used to calculate CCT during deformation using a ray-tracing method to correct for Scheimpflug and optical distortion within each image. Correlation and stepwise regression analyses between the shape-related DCR parameters and %ΔCCT were conducted. A mixed model analysis was performed to test the effect of IOP and the strongest significant predictors of the stepwise regression on %ΔCCT. The significance threshold was set to p < 0.05.
Results:
Thirty eyes were ultimately analyzed and CCT increased significantly from the pre-deformation state to the highest concavity state at each IOP level (p < 0.001). IOP and multiple shape DCRs were found to be significantly related to %ΔCCT (p < 0.0001). The strongest predictor of %ΔCCT was integrated inverse radius (IIR) (p < 0.0001; partial R2 = 0.4772) with no other parameter having a partial R2 value greater than 0.04. The mixed model analysis showed that IIR was the sole predictor (p = 0.0098) and IOP was no longer significant as a single predictor. However, the interaction of IIR with IOP (p = 0.0023) had a significant effect on %ΔCCT.
Conclusion:
Percent change in CCT is influenced by corneal stiffness as indicated by the significant relationship with IIR. The %ΔCCT may be a potential biomarker for determining differences in corneal deformation response with corneal diseases.
Related Concept Videos
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...
Deformations in a Symmetric Member in Bending
When the member is segmented into tiny cubic elements, it is observed that the primary stress...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Temperature Dependent Deformation
Plastic Deformation in Circular Shafts