Dosimetric study for breathing-induced motion effects in an abdominal pancreas phantom for carbon ion mini-beam
Christina Stengl1,2,3, Iván D Muñoz2,3,4, Eric Arbes4,5
1Medical Faculty Heidelberg, Heidelberg University, Heidelberg, Germany.
Background:
Particle mini-beam therapy exhibits promise in sparing healthy tissue through spatial fractionation, particularly notable for heavy ions, further enhancing the already favorable differential biological effectiveness at both target and entrance regions. However, breathing-induced organ motion affects particle mini-beam irradiation schemes since the organ displacements exceed the mini-beam structure dimensions, decreasing the advantages of spatial fractionation.
Purpose:
In this study, the impact of breathing-induced organ motion on the dose distribution was examined at the target and organs at risk(OARs) during carbon ion mini-beam irradiation for pancreatic cancer.
Methods:
As a first step, the carbon ion mini-beam pattern was characterized with Monte Carlo simulations. To analyze the impact of breathing-induced organ motion on the dose distribution of a virtual pancreas tumor as target and related OARs, the anthropomorphic Pancreas Phantom for Ion beam Therapy (PPIeT) was irradiated with carbon ions. A mini-beam collimator was used to deliver a spatially fractionated dose distribution. During irradiation, varying breathing motion amplitudes were induced, ranging from 5 to 15 mm. Post-irradiation, the 2D dose pattern was analyzed, focusing on the full width at half maximum (FWHM), center-to-center distance (ctc), and the peak-to-valley dose ratio (PVDR).
Results:
The mini-beam pattern was visible within OARs, while in the virtual pancreas tumor a more homogeneous dose distribution was achieved. Applied motion affected the mini-beam pattern within the kidney, one of the OARs, reducing the PVDR from 3.78 0.12 to 1.478 0.070 for the 15 mm motion amplitude. In the immobile OARs including the spine and the skin at the back, the PVDR did not change within 3.4% comparing reference and motion conditions.
Conclusions:
This study provides an initial understanding of how breathing-induced organ motion affects spatial fractionation during carbon ion irradiation, using an anthropomorphic phantom. A decrease in the PVDR was observed in the right kidney when breathing-induced motion was applied, potentially increasing the risk of damage to OARs. Therefore, further studies are needed to explore the clinical viability of mini-beam radiotherapy with carbon ions when irradiating abdominal regions.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...