Severe Hemodynamic Collapse During Humerus Stabilization with Photodynamic Implant: A Report of Two Cases
Samir Sabharwal1, Patrick J Boland, Max Vaynrub
1Department of Surgery, Memorial Sloan Kettering Cancer Center, New York, New York.
Case:
We present 2 cases of severe hemodynamic collapse during prophylactic stabilization of impending pathologic humerus fractures using a photodynamic bone stabilization device. Both events occurred when the monomer was infused under pressure into a balloon catheter.
Conclusion:
We suspect that an increase in intramedullary pressure during balloon expansion may cause adverse systemic effects similar to fat embolism or bone cement implantation syndrome. Appropriate communication with the anesthesia team, invasive hemodynamic monitoring, and prophylactic vent hole creation may help mitigate or manage these adverse systemic effects.