Zefeng Wang1, Zhongyu Yuan1, Haiwei Li1
1Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China.
Abstract:
We report three patients with screw-in lead perforation in the right atrial free wall not long after device implantation. All the patients complained of intermittent stabbing chest pain associated with deep breathing during the implantation. The "dry" epicardial puncture was utilized to avoid hemopericardium during lead extraction in the first case. The atrial electrode was repositioned in all cases and replaced by a new passive fixation lead in two patients with resolution of the pneumothorax or pericardial effusion. A literature review of 50 reported cases of atrial lead perforation was added to the findings in our case report.