Background:
Standardized and validated heat inactivation procedure for Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) are not available. For heat inactivation, various protocols were reported to prepare External Quality Assessment Programme (EQAP) samples without direct comparison between different durations.
Objective:
To assess the heat inactivation procedures against SARS-CoV-2. The efficacy of the optimized condition was reflected by the results from laboratories testing the EQAP samples.
Study Design:
The SARS-CoV-2 strain was exposed to 95 °C in a water bath for three different time intervals, 5 min, 10 min and 15 min, respectively. The efficacy of inactivation was confirmed by the absence of cytopathic effects and decreasing viral load in 3 successive cell line passages. The viral stock inactivated by the optimal time interval was dispatched to EQAP participants and the result returned were analyzed.
Results:
All of the three conditions were capable of inactivating the SARS-CoV-2 of viral load at around cycle threshold value of 10. When the 95 °C 10 min condition was chosen to prepare SARS-CoV-2 EQAP samples, they showed sufficient homogeneity and stability. High degree of consensus was observed among EQAP participants in all samples dispatched.
Conclusions:
The conditions evaluated in the present study could be helpful for laboratories in preparing SARS-CoV-2 EQAP samples.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...