Related Experiment Video
Updated: Jun 28, 2025

Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Carboxyhemoglobin predicts oxygenator performance and imminent oxygenator change in extracorporeal membrane
Rolf Erlebach1, Alix Buhlmann1, Rea Andermatt1
1Institute of Intensive Care Medicine, University Hospital Zurich, Zurich, Switzerland.
Background:
The continuous exposure of blood to a non-biological surface during extracorporeal membrane oxygenation (ECMO) may lead to progressive thrombus formation in the oxygenator, hemolysis and consequently impaired gas exchange. In most centers oxygenator performance is monitored only on a once daily basis. Carboxyhemoglobin (COHb) is generated upon red cell lysis and is routinely measured with any co-oximetry performed to surveille gas exchange and acid-base homeostasis every couple of hours. This retrospective cohort study aims to evaluate COHb in the arterial blood gas as a novel marker of oxygenator dysfunction and its predictive value for imminent oxygenator change.
Results:
Out of the 484 screened patients on ECMO 89, cumulatively requiring 116 oxygenator changes within 1833 patient days, including 19,692 arterial COHb measurements were analyzed. Higher COHb levels were associated with lower post-oxygenator pO2 (estimate for log(COHb): - 2.176 [95% CI - 2.927, - 1.427], p < 0.0001) and with a shorter time to oxygenator change (estimate for log(COHb): - 67.895 [95% CI - 74.209, - 61.542] hours, p < 0.0001). COHb was predictive of oxygenator change within 6 h (estimate for log(COHb): 5.027 [95% CI 1.670, 15.126], p = 0.004).
Conclusion:
COHb correlates with oxygenator performance and can be predictive of imminent oxygenator change. Therefore, longitudinal measurements of COHb in clinical routine might be a cheap and more granular candidate for ECMO surveillance that should be further analyzed in a controlled prospective trial design.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Transport in the Blood
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...