Quality control in the Netherlands; todays practices and starting points for guidance and future research
Huub H van Rossum1, Marith van Schrojenstein Lantman2,3,4, Michel Severens5
1Department of Laboratory Medicine, Netherlands Cancer Institute, Amsterdam, The Netherlands.
Objectives:
Adequate analytical quality of reported results is primarily ensured by performing internal quality control (iQC). Currently, several different iQC practices are in use. As a prelude to the revision of a Dutch guidance document on analytical QC, a questionnaire was sent out to gain insights in the applied practices and the need for guidance.
Methods:
A questionnaire, containing 20 multiple-choice questions with possibilities for explanation and comment on iQC practices and aspects was distributed to all clinical chemistry laboratories within the Netherlands. Results were reported descriptively.
Results:
Responses were received from 27 clinical laboratories (response 43 %). In 30 % the iQC was based on the analytical characteristics only, while 30 % used a 6-Sigma method, 19 % risk-based beyond 6-Sigma and 22 % used an alternative approach. 89 % of laboratories used a virtual analyzer model for iQC setup within one or more laboratory sites. Practices for determining standard deviation (SD) values included determining SD for each new iQC material (35 %), using historical SD values for new materials (35 %), and incorporating clinical tolerances into the SD value (31 %). Furthermore, 44 % of laboratories used patient moving averages for one or more tests. Daily iQC management was based on either "traffic lights" indicating in or out of control status, and review of all QC charts, often using multiple software systems.
Conclusions:
A large heterogeneity of iQC practices in clinical laboratories was observed in the Netherlands. Several starting points for further research and/or guidance were identified, particularly in relation to the determination of SD values, the virtual analyzer model and methods to ensure analyzer equivalence.
More Related Videos
08:53Safety Precautions and Operating Procedures in an ABSL-4 Laboratory: 2. General Practices
Published on: October 3, 2016
08:52Improving Reproducibility to Meet Minimal Information for Studies of Extracellular Vesicles 2018 Guidelines in Nanoparticle Tracking Analysis
Published on: November 17, 2021
Related Concept Videos
Quality Assurance
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Data Validation
Key parameters for method validation include:
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Methods of Documentation V: CBE
In CBE, healthcare professionals establish predefined standards of practice that define what constitutes...