Derivation of Stroke Volume from Pulmonary Artery Pressures
Ivan H W Yim1,2, Nigel E Drury1,2, Hoong Sern Lim3,4
1University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK.
Introduction:
Intermittent cardiac output (CO) studies using thermodilution are considered the gold standard. We have developed a stroke volume (SV) calculator from pulmonary pulse pressure (PP) to allow continuous monitoring of SV and CO from PP.
Methods:
Hemodynamic data on 169 patients following orthotopic heart transplantation were used to compare our calculator-derived SV (and SV index, or SVi) against thermodilution-derived SV on admission into intensive care unit immediately following heart transplantation (T0) and 6 h after admission (T6).
Results:
The calculated SV correlated with thermodilution-derived SV T0 (r = 0.920, p < 0.001, coefficient of 0.539 and the constant of 2.06). The median calculator SV, adjusted for coefficient and constant, was 48.4 ml (37.7, 60.7), comparable to the median thermodilution-derived SV 47.9 ml (37.5, 61.0), p = 0.737 with acceptable agreement on Bland-Altman plots. The thermodilution-derived SVi was 28.1 ml (19.7, 38.7) and adjusted calculator-derived SVi 28.9 ml (19.7, 39.9), p = 0.781. At T6, median thermodilution-derived SVi was 27.7 ml (19.5, 35.9) compared to the calculator-derived SVi median of 26.1 ml (17.7, 37.7), p = 0.203.
Conclusions:
Changes in PP can be used to track changes in SV using this calculator. Changes in PP may be used to assess response to treatment in the early post-operative period.
Related Concept Videos
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Vascular Resistance
The primary determinants of vascular resistance are vessel diameter, blood viscosity, and vessel length. Among these, vessel diameter plays the most significant role due to the fourth power relationship described by...
Bernoulli's Equation: Problem Solving
The first step is to compute the cross-sectional areas of the pipe and the Venturi throat to analyze the pressure difference indicated by the pressure gauge. Next, the continuity...