Two cases with discrepancy in the quantitative cytological assessment of cerebrospinal fluid in neonatal samples
Pavel Broz1, Simona Kukralova1, Jana Palatova1
1Institute of Clinical Biochemistry and Haematology, University Hospital in Pilsen, Pilsen, Czech Republic.
Abstract:
We present two cases from the neonatal department with cerebrospinal fluid examination. We revealed a striking discrepancy in polymorphonuclear (PMN) and mononuclear (MN) cell counts using conventional light microscopy in comparison with automated analyzer Sysmex XN-1000 (PMNs - 13 vs. 173x106/L, MNs - 200 vs. 67x106/L in case 1 and PMNs - 13 vs. 372x106/L, MNs - 411 vs. 179x106/L in case 2). We revealed the dominant presence of hemosiderophages in both cases in cytospin slide. Even though Sysmex XN-1000 offers fast examination with a low sample volume, there is possibility of misdiagnosis, with negative impact on the patient.