Background:
With the widespread use of low-dose spiral computed tomography (LDCT) and increasing awareness of personal health, the detection rate of pulmonary nodules is steadily rising.
Objective:
To evaluate the success rate and safety of two different models of Hook-Wire needle localization procedures for pulmonary small nodule biopsy.
Methods:
Ninety-four cases with a total of 97 pulmonary small nodules undergoing needle localization biopsy were retrospectively analyzed. The cases were divided into two groups: Group A, using breast localization needle steel wire (Bard Healthcare Science Co., Ltd.); Group B, using disposable pulmonary nodule puncture needle (SensCure Biotechnology Co., Ltd.). All patients underwent video-assisted thoracoscopic surgery (VATS) for nodule removal on the same day after localization and biopsy. The puncture localization operation time, success rate, complications such as pulmonary hemorrhage, pneumothorax, hemoptysis, and postoperative comfort were observed and compared.
Results:
In Group A, the average localization operation time for 97 nodules was 15.47 ± 5.31 minutes, with a success rate of 94.34%. The complication rate was 71.69% (12 cases of pneumothorax, 35 cases of pulmonary hemorrhage, 2 cases of hemoptysis), and 40 cases of post-localization discomfort were reported. In Group B, the average localization operation time was 25.32 ± 7.83 minutes, with a 100% success rate. The complication rate was 29.55% (3 cases of pneumothorax, 15 cases of pulmonary hemorrhage, 0 cases of hemoptysis), and 3 cases reported postoperative discomfort. According to the data analysis in this study, Group B had a lower incidence of puncture-related complications than Group A, along with a higher success rate and significantly greater postoperative comfort.
Conclusions:
The disposable pulmonary nodule puncture needle is safer and more effective in pulmonary small nodule localization biopsy, exhibiting increased comfort compared to the breast localization needle. Additionally, the incidence of complications is significantly lower.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...