On the Feasibility of Using Behavioral Listening Effort Test Methods to Evaluate Auditory Performance in Cochlear
Maartje M E Hendrikse1, Gertjan Dingemanse1, André Goedegebure1
1Department of Otorhinolaryngology and Head and Neck Surgery, Erasmus MC University Medical Center, Rotterdam, the Netherlands.
Trends in Hearing
|April 27, 2024
Summary
Two listening effort tests for cochlear implant (CI) users showed significant differences between signal-to-noise ratios but were not sensitive enough for individual clinical use. These tests are suitable for group research data analysis.
Area of Science:
- Audiology
- Neuroscience
- Rehabilitation Engineering
Background:
- Assessing hearing aid and cochlear implant (CI) fitting requires realistic outcome measures for everyday hearing challenges.
- Listening effort measures may offer greater sensitivity than speech intelligibility measures for detecting differences in hearing-device settings when intelligibility is near maximal.
Purpose of the Study:
- To investigate the feasibility of two tests for measuring changes in listening effort in CI users due to signal-to-noise ratio (SNR) differences.
- To evaluate the suitability of these tests for clinical use by comparing the effect size of SNR differences on listening effort measures with test-retest differences.
Main Methods:
- Nineteen CI users participated in two listening effort tests (sentence-final word identification and recall test [SWIRT] and sentence verification test [SVT]) at two SNRs (+4 and +8 dB).
- Dual-task paradigms were employed to assess listening effort under varying acoustic conditions.
Main Results:
- Both SWIRT and SVT demonstrated a significant difference in listening effort between the tested SNRs.
- However, the effect size of these differences was comparable to test-retest variability.
- The sensitivity of these measures was not superior to traditional speech intelligibility measures.
Conclusions:
- The implemented SWIRT and SVT are not suitable for clinical measurement of listening effort differences in individual CI users.
- These tests can be effectively utilized in research settings for analyzing group data from CI users.