Effect of increased resistance on dynamic compliance assessed by two clinical monitors during volume-controlled
Daniel Ben-Aderet1, João Hn Soares2, Melissa Lc Bueno2
1William Pritchard Veterinary Medical Teaching Hospital, School of Veterinary Medicine, University of California Davis, Davis, CA, USA.
Objective:
To evaluate the effect of increased respiratory system resistance (RRS) on dynamic compliance (Cdyn) assessed by the NM3 monitor (Cdyn(NM3)) and the E-CAiOV module (Cdyn(ECAiOV)).
Study Design:
Prospective laboratory study.
Methods:
A training test lung (TTL) simulated the mechanical ventilation of a mammal with 50 and 300 mL tidal volumes in three conditions of RRS [normal (RBL), moderately increased (R1) and severely increased (R2)] and a wide range of clinically relevant Cdyn. Simulations at increased RRS were paired with simulations at RBL with the same static compliance for comparisons. Pearson's correlation coefficient and concordance correlation coefficient between the measurements at RBL with the ones with increased RRS were calculated. Bland-Altman plots were also used to evaluate the agreement of Cdyn(ECAiOV) and Cdyn(NM3) at RBL (control values) with their paired values at R1 and R2. Relative bias and limits of agreement (LOAs) were calculated and LOAs larger than 30% were considered unacceptable. Trending ability of Cdyn(NM3) and Cdyn(ECAiOV) were evaluated by polar plots. Values of p < 0.05 were considered significant.
Results:
The effect of increased RRS was more pronounced for Cdyn(ECAiOV) than for Cdyn(NM3). Unacceptable agreement was only observed in Cdyn(NM3) at R2 in the 300 mL simulation (bias = -18.3% and lower LOA = -45%). For Cdyn(ECAiOV), agreement was unacceptable for all tested RRS in both simulations, being the worst at R2 in the 300 mL simulation (bias = -54.7% and lower LOA = -100.2%). Both levels of increased RRS caused poor trending ability for Cdyn(ECAiOV), whereas the same effect was only observed for Cdyn(NM3) at R2.
Conclusions And Clinical Relevance:
In the presence of increased RRS, Cdyn estimated by the NM3 monitor presented better capability to distinguish between changes in RRS from changes in respiratory system compliance.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Breathing
Lung Capacity