Dosimetric Systems in Pre-Treatment QA for Stereotactic Treatments: Correlation Agreements and Target Volume
Sivakumar Muthu1,2, Gopinath Mudhana1
1Department of Physics, School of Advanced Sciences, Vellore Institute of Technology, Chennai - 600 127, India.
Aim:
This study comprehensively investigated pre-treatment quality assurance (QA) for 100 cancer patients undergoing stereotactic treatments (SRS/SRT) using various detectors.
Methods:
The study conducted QA for SRS/SRT treatments planned with a 6MV SRS beam at a dose rate of 1,000 MU/min, utilizing Eclipse v13.6 Treatment Planning System (TPS). Point dose measurements employed 0.01cm3 and 0.13cm3 cylindrical ionization chambers, while planar dose verification utilized Gafchromic EBT-XD Film and Portal Imager (aS1000). Plans were categorized by target volume, and a thorough analysis compared point dose agreements, planar dose gamma pass rates, and their correlations with chamber volume mean dose, detector type, and point dose agreement. Additionally, the consistency between different ionization chambers was assessed.
Results:
Point dose agreement generally improved with increasing target volume, except for volumes over 10cm3 with 0.01cm3 chambers, showing a contrary trend. Significant differences (p<0.05) were observed between TPS and measured doses for both chambers. Gamma pass rate improved with increasing target volume in EBT XD and aS1000 analyses, except for the >10cm3 group in EBT XD. EBT XD demonstrated better agreement with TPS for target volumes up to 10cm3 compared to aS1000, with a statistically significant difference (p<0.05) between the detectors. Strong correlations were found between chamber point dose and chamber volume mean dose agreement, as well as between the two gamma criteria analyses of the same detector type in the planar dose correlation analysis. However, weak correlations were discovered for other analyses.
Conclusion:
This study found weak correlation between different detector types in pre-treatment QA for point dose and planar dose evaluation. However, within a specific detector type, strong correlation was observed for different point dose evaluation methods and gamma criteria. This highlights the importance of cautious interpretation of QA results, particularly for SRS QA, due to the lack of correlation between detector types.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Data Validation
Key parameters for method validation include:
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Measurement of Bioavailability: Pharmacodynamic Methods
Dosage Regimen: Individualization
Therapeutic Drug Monitoring: Drug Analysis Methods