Related Experiment Video
Updated: Jun 27, 2025

Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
Establishment and Validation of an Automated System for the Antifactor IIa Assay: A Case Study of Potency Assessment
Gokselin Ozgen1, Merve Turk Gezer1, Guliz Armagan2
1Abdi İbrahim Pharmaceuticals Research & Development Center (R&D), Istanbul 34538, Türkiye.
Abstract:
Antithrombotic agents and anticoagulant drugs, such as those from the heparin family, are employed in clinical settings for the prevention and treatment of clotting, thromboembolism, and wound healing. The potency assessment of antithrombotic agents is typically conducted using antifactor IIa assay with manual systems which are time-consuming and often lack repeatability. Here, we present a novel automated system that significantly enhances assay repeatability, attaining an outstandingly low relative standard deviation (RSD) % of only 0.6% for repeatability. This system has been applied to a pharmaceutical gel formulation for wound healing developed by Abdi Ibrahim Pharmaceuticals R&D Center as a case study for validation. The automated system demonstrated substantial improvements over manual systems in linearity (R2 = 0.9927), precision, accuracy, specificity, and robustness. The system aligns with the European Pharmacopoeia specifications, promising to enhance quality control across pharmaceutical formulations and conduct absorbance-based end-point assays within the pharmaceutical industry while offering increased throughput and cost-effectiveness.