Comparative analysis of dimensional changes in autoclavable polyvinyl siloxane (PVS) impressions under various
Noha Taymour1, Sanaa Hussein Abdel Kader2, Moustafa N Aboushelib3
1Department of Substitutive Dental Sciences, College of Dentistry, Imam Abdulrahman Bin Faisal University, P.O. Box 1982, Dammam 31441, Saudi Arabia.
Background:
Impressions that maintain their dimensional stability after autoclaving effectively control cross-infection and contamination resulting from a patient's oral secretions.
Purpose:
The study aimed to assess the dimensional stability of autoclavable polyvinyl siloxanes after disinfection and sterilization.
Methods:
A stainless steel metal model containing three full veneer crown preparations was fabricated according to ANSI/ADA specification No. 19. Reference grooves were established on the occlusal and axial surfaces of the abutments for accurate measurements. Forty impressions were created from the master model using single-step impression technique monophase polyvinyl siloxane material (AFFINIS, Coltene/Whaledent, Altstatten, Switzerland). The impressions were categorized into four groups: Group A (control, ten untreated impressions), Group B (ten disinfected impressions with 5.25 % sodium hypochlorite [NaOCl]), Group C (ten disinfected impressions with 2 % glutaraldehyde), and Group D (ten autoclaved impressions at 134 °C for 18 min). Subsequently, stone casts were produced using type IV gypsum products (Gelstone R, BK Giulini Chemie, Ludwigshafen/Rh., Germany). The dimensional accuracy of the obtained casts was assessed by measuring the inter-abutment measurements (between the abutments) and the intra-abutment measurements (diameter and height of the abutments). These measurements were performed using a universal measuring microscope (Olympus stereomicroscope B061 Imaging Corp. Tokyo, Japan) with a precision of 0.001 mm. The dimensions of the stone casts from the study groups were then compared to those of the control group. Data analysis was performed using a one-way ANOVA with a significance level of α = 0.05.
Results:
AFFINIS impressions subjected to chemical disinfection in 5.25 % NaOCl and 2 % glutaraldehyde with different immersion times showed slight expansion in the intra- and inter-abutment measurements. The impressions autoclaved at 134 °C for 18 min showed slight shrinkage in the intra- and inter-abutment measurements. The dimensional change was statistically non-significant, and the percent of dimensional changes within the experimental groups was within the clinically accepted limit (α < 0.5).
Conclusion:
AFFINIS polyvinyl siloxanes retain dimensional stability suitable for clinical use when subjected to chemical disinfection and steam autoclaving.
More Related Videos
05:30Roughness Impact of Piezoelectric Dental Scaler on Two Distinct Flowable Composite Filling Materials
Published on: January 10, 2025
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...