Related Experiment Video
Updated: Jun 27, 2025

A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
An easy method to determine crucial AMEI performance parameters from clinical routine data in individuals - Part 1:
Hannes Maier1,2, Thomas Lenarz1,2, Susan Busch1,2
1Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
Objective:
The frequency specific maximum output (MO) of active middle ear implants is the most crucial parameter for speech intelligibility. We determined individual MO from clinical routine data in round window (RW) coupling of the Vibrant Soundbridge (VSB).
Design:
Monocentric, retrospective analysis.
Study Sample:
68 ears implanted with the VSB at the RW were analysed. Using bone conduction and direct threshold, MO was determined for combinations of implants (VORP502, VORP503) and processors (Samba, Amadé). Coupling modes were: (A) without coupler (N = 28), (B) spherical coupler (N = 19), (C) soft coupler (N = 10) or (D) custom-made "Hannover coupler" (N = 11).
Results:
The MO frequency dependence was similar for coupling types (A-D) with a maximum at 1.5 kHz. No differences between groups were observed, although the average MO of the soft coupler was 10 dB lower. The average MO (0.5, 1.0, 2.0, 4.0 kHz) was (A) 77.6 ± 15.0 dB HL, (B) 81.0 ± 11.1 dB HL, (C) 67.6 ± 17.9 dB HL (C), and (D) 79.6 ± 11.7 dB HL (D).
Conclusion:
The individual MO can be determined from patients' clinical data. It permits in-depth analyses of patient outcomes and definition of evidence-based indication and decision criteria.
More Related Videos
00:08A Cross-Disciplinary and Multi-Modal Experimental Design for Studying Near-Real-Time Authentic Examination Experiences
Published on: September 4, 2019
04:05Author Spotlight: Unveiling Prognostic Indicators in Heart Failure - The Role of Phase Angle and Bioelectrical Impedance Analysis
Published on: June 30, 2023
Related Concept Videos
Kaplan-Meier Approach
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.