Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Diagnostic accuracy for detecting atrial fibrillation using a novel machine learning algorithm in a blood pressure
Matthew Janik1, George Raad2, George Nijmeh3
1Cardiology Division, Wilmington Health, Wilmington, North Carolina.
Background:
Early detection of atrial fibrillation (AF) is key for preventing strokes. Blood pressure monitors (BPMs) with built-in AF screening features have the potential for early detection at home. Recently, 2 BPMs (HEM-7371T1-AZ and HEM-7372T1-AZAZ, Omron Healthcare Co., Ltd.) that share a novel AF screening feature have been developed. Their AF screening feature utilizes an algorithm that incorporates machine learning, with the potential to improve diagnostic accuracy.
Objective:
The purpose of this study was to evaluate the performance of this AF screening feature in a multicenter, prospective clinical study at 5 sites in the United States.
Methods:
A total of 559 subjects were enrolled for this study: 267 in AF cohort and 292 in the non-AF cohort. AF screening was performed in all subjects by the 2 Omron BPMs and by 1 Microlife BPM (BP 3MX1-3, WatchBP Home A, Microlife Corp.), and a simultaneous 12-lead electrocardiogram (ECG) was recorded for comparison. All 12-lead ECGs were interpreted by a board-certified cardiologist who was blinded to the BPM results. Sensitivity, specificity, and accuracy for the diagnosis of AF were calculated.
Results:
Omron HEM-7371T1-AZ BPM had sensitivity of 95.1% (95% confidence interval [CI] 91.8%-97.4%), specificity 98.6% (95% CI 96.6%-99.7%), and accuracy of 97.0% (95% CI 95.2%-98.2%). Equivalent results were obtained with the Omron HEM-7371T1-AZAZ BPM. This compared favorably to the Microlife BPM (sensitivity 78.5%, 95% CI 73.1%-83.3%; specificity 97.6%, 95% CI 95.1%-99.0%; accuracy 88.4%, 95% CI 85.5%-91.0%).
Conclusion:
These data support both home and professional use of these novel Omron BPMs for the detection of AF.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...