Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Pulse rhythm01:30

Pulse rhythm

787
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
787
Hypoglycemia and Glucagon01:15

Hypoglycemia and Glucagon

257
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
257
Glucose Homeostasis: Regulation of Blood Glucose01:02

Glucose Homeostasis: Regulation of Blood Glucose

1.6K
Carbohydrates consumed through foods are converted into glucose, a crucial energy source for the body. In the prandial state, high blood glucose levels stimulate the secretion of insulin from the pancreas. Insulin inhibits hepatic glucose production and stimulates glucose uptake and metabolism by muscle and adipose tissue. The excess glucose is converted into glycogen and stored in the liver and muscles.
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
1.6K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Building a scalable, multilevel healthcare pipeline program through structured mentoring families: perceptions and outcomes from two cohorts.

Journal of the National Medical Association·2026
Same author

Do Improving Hepatic Biomarkers Predict Spontaneous Stone Passage? A Retrospective Analysis of ERCP Yield in Suspected Choledocholithiasis.

Digestive diseases and sciences·2026
Same author

Management of bone health in adult solid organ transplant recipients: a multidisciplinary care pathway.

Clinical transplantation and research·2026
Same author

Symptomatic Benign Prostate Hyperplasia May Increase the Risk of Kidney Stones.

Journal of endourology·2026
Same author

Vedolizumab for stricturing Crohn's disease of the terminal ileum - an international multicenter cohort study.

BMC gastroenterology·2026
Same author

Continuous Glucose Monitoring for Postkidney Transplantation in Patient With Pre-existing Type 2 Diabetes.

Endocrine practice : official journal of the American College of Endocrinology and the American Association of Clinical Endocrinologists·2026

Related Experiment Video

Updated: Jun 27, 2025

Simple Continuous Glucose Monitoring in Freely Moving Mice
03:25

Simple Continuous Glucose Monitoring in Freely Moving Mice

Published on: February 24, 2023

5.4K

Continuous Glucose Monitoring Using the Dexcom G6 in Cardiac Surgery During the Postoperative Period.

Steven R Insler1, Brett Wakefield1, Andrea Debs2

  • 1Departments of Intensive Care and Resuscitation and Cardiothoracic Anesthesiology, The Cleveland Clinic Foundation, Cleveland, Ohio.

Endocrine Practice : Official Journal of the American College of Endocrinology and the American Association of Clinical Endocrinologists
|May 1, 2024
PubMed
Summary

This pilot study assessed the accuracy of the Dexcom G6Pro continuous glucose monitor (CGM) in postsurgical cardiac patients. While most readings were clinically acceptable, the device

Keywords:
Arterial Blood GasCardiopulmonary BypassCardiovascular Intensive Care UnitContinuous Glucose monitorGlucoseHyperglycemia

More Related Videos

Improving IV Insulin Administration in a Community Hospital
12:08

Improving IV Insulin Administration in a Community Hospital

Published on: June 11, 2012

18.8K
Surgical Tips and Tricks for Performing Porcine Pancreas Transplantation
10:32

Surgical Tips and Tricks for Performing Porcine Pancreas Transplantation

Published on: July 20, 2022

1.8K

Related Experiment Videos

Last Updated: Jun 27, 2025

Simple Continuous Glucose Monitoring in Freely Moving Mice
03:25

Simple Continuous Glucose Monitoring in Freely Moving Mice

Published on: February 24, 2023

5.4K
Improving IV Insulin Administration in a Community Hospital
12:08

Improving IV Insulin Administration in a Community Hospital

Published on: June 11, 2012

18.8K
Surgical Tips and Tricks for Performing Porcine Pancreas Transplantation
10:32

Surgical Tips and Tricks for Performing Porcine Pancreas Transplantation

Published on: July 20, 2022

1.8K

Area of Science:

  • Cardiology
  • Endocrinology
  • Medical Device Technology

Background:

  • Cardiac surgery frequently causes hyperglycemia, linked to poor patient outcomes like infections and mortality.
  • Effective glucose monitoring is crucial for managing postsurgical hyperglycemia and improving patient recovery.

Purpose of the Study:

  • To evaluate the accuracy of the Dexcom G6Pro continuous glucose monitor (CGM) compared to standard methods in the immediate postoperative period after cardiac surgery.
  • To assess CGM performance amidst common postsurgical complications such as fluid shifts and inflammatory responses.

Main Methods:

  • A prospective study involving 29 patients undergoing cardiac surgery.
  • Correlation analysis between 1428 clinical glucose readings and 45,645 Dexcom G6Pro CGM data points.
  • Data analysis included Clarke and consensus error grids, with subsets excluding the initial 24 hours post-ICU admission.

Main Results:

  • Over 99% of Dexcom G6Pro CGM readings fell into clinically acceptable zones (A and B) on error grids.
  • Mean absolute relative difference between CGM and clinical measurements was approximately 20-22%.
  • No adverse events were reported during the study period.

Conclusions:

  • The Dexcom G6Pro CGM demonstrated a high rate of clinically acceptable glucose readings in postsurgical cardiac patients.
  • Despite acceptable clinical decision-making, the mean absolute relative difference limits its current use for direct treatment decisions.
  • Continuous glucose monitoring holds future potential for managing glycemic fluctuations and improving patient outcomes in critical care settings.