Six-month electrical performance of the first dual-chamber leadless pacemaker
Gerhard Hindricks1, Rahul Doshi2, Pascal Defaye3
1Deutsches Herzzentrum der Charite, Berlin, Germany.
Heart Rhythm
|May 2, 2024
Summary
The first dual-chamber leadless pacemaker system shows reliable electrical performance over six months. This innovative device offers stable pacing and sensing for patients needing synchronized heart rhythm management.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- The novel dual-chamber leadless pacemaker (DC-LP) system utilizes two independent atrial and ventricular devices for synchronized pacing.
- Previous reports established the initial safety and efficacy of this groundbreaking technology.
Purpose of the Study:
- To assess the long-term electrical performance and stability of the dual-chamber leadless pacemaker (DC-LP) system.
- Evaluate the chronic electrical characteristics of the leadless pacemaker system over a six-month period.
Main Methods:
- Implantation of the DC-LP system (Aveir DR, Abbott) in patients requiring dual-chamber pacing.
- Monitoring of pacing capture thresholds, sensed amplitudes, and pacing impedance at regular intervals from implant to six months.
Main Results:
- Significant improvements in pacing capture thresholds and sensed amplitudes for both atrial and ventricular leadless pacemakers within the first month.
- Stable pacing impedances in atrial leadless pacemakers and a reduction in ventricular leadless pacemaker impedances.
- Electrical parameters remained consistent from one to six months post-implantation, irrespective of implant location.
Conclusions:
- The dual-chamber leadless pacemaker system demonstrated dependable electrical performance during the initial six-month evaluation.
- This study confirms the reliable chronic electrical function of the first dual-chamber leadless pacemaker system in human subjects.