Optimum Parameters in Ultrasound Coherent Plane Wave Compounding for High LPWV Estimation: Validation on Phantom and
Li Xiong1, Yufeng Zhang1, Bingbing He1
1Department of Electronic Engineering, Information School, Yunnan University, Kunming, China.
Objectives:
The aim of this study is to determine the optimum and fine values of the number and transmission angles of tilted plane waves for coherent plane-wave compounding (CPWC)-based high local pulse wave velocity (LPWV) estimation.
Methods:
A Verasonics system incorporating a linear array probe L14-5/38 with 128 elements and a pulsatile pump, CompuFlow1000, were used to acquire radio frequency data of 3, 5, 7, and 9 tilted plane wave sequences with angle intervals from 0° to 12° with a coarse interval increment step of 1°, and the angle intervals from 0° to 2° with a fine interval increment step of 0.25° from a carotid vessel phantom with the LPWV of 13.42 ± 0.90 m/s. The mean value, standard deviation, and coefficients of variation (CV) of the estimated LPWVs were calculated to quantitatively assess the performance of different configurations for CPWC-based LPWV estimation. Ten healthy human subjects of two age groups were recruited to assess the in vivo feasibility of the optimum parameter values.
Results:
The CPWC technique with three plane waves (PRF of 12 kHz corresponding to a frame rate of 4000 Hz) with an interval of 0.75° had LPWVs of 13.52 ± 0.08 m/s with the lowest CV of 1.84% on the phantom, and 5.49 ± 1.46 m/s with the lowest CV of 12.35% on 10 subjects.
Conclusions:
The optimum parameters determined in this study show the best repeatability of the LPWV measurements with a vessel phantom and 10 healthy subjects, which support further studies on larger datasets for potential applications.
More Related Videos
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Wave Parameters
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
Upsampling
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....
Linear Approximation in Time Domain
For a simple pendulum with a mass evenly distributed along its length and the center of mass located at half the pendulum's length,...