Related Experiment Video
Updated: Jun 27, 2025

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Freehand 3D Ultrasound Imaging Based on Probe-mounted Vision and IMU System
Weizhen He1, Bingshuai Zhao1, Yongjin Zhou2
1School of Biomedical Engineering, Shenzhen University Medical School, Shenzhen University, Shenzhen, China.
Objectives:
Freehand three-dimensional (3D) ultrasound (US) is of great significance for clinical diagnosis and treatment, it is often achieved with the aid of external devices (optical and/or electromagnetic, etc.) that monitor the location and orientation of the US probe. However, this external monitoring is often impacted by imaging environment such as optical occlusions and/or electromagnetic (EM) interference.
Methods:
To address the above issues, we integrated a binocular camera and an inertial measurement unit (IMU) on a US probe. Subsequently, we built a tight coupling model utilizing the unscented Kalman algorithm based on Lie groups (UKF-LG), combining vision and inertial information to infer the probe's movement, through which the position and orientation of the US image frame are calculated. Finally, the volume data was reconstructed with the voxel-based hole-filling method.
Results:
The experiments including calibration experiments, tracking performance evaluation, phantom scans, and real scenarios scans have been conducted. The results show that the proposed system achieved the accumulated frame position error of 3.78 mm and the orientation error of 0.36° and reconstructed 3D US images with high quality in both phantom and real scenarios.
Conclusions:
The proposed method has been demonstrated to enhance the robustness and effectiveness of freehand 3D US. Follow-up research will focus on improving the accuracy and stability of multi-sensor fusion to make the system more practical in clinical environments.