Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
Comparison of two electronic dynamometers for measuring handgrip strength
Sebastian Keller Rolsted1, Kasper Dyrmose Andersen1, Gustav Dandanell1
1Department of Physical and Occupational Therapy, Copenhagen University Hospital, Bispebjerg and Frederiksberg, Copenhagen, Denmark; Centre for Health and Rehabilitation, University College Absalon, Slagelse, Denmark.
Handgrip strength testing results from the Baseline BIMS Digital Grip Dynamometer and Jamar+ are interchangeable. This study found excellent agreement and criterion validity between the two devices, indicating low measurement error for reliable physical condition assessments.
Area of Science:
- Biomedical Engineering
- Kinesiology
- Gerontology
Background:
- Handgrip strength (HGS) is a key indicator of overall physical health and function.
- Standardized HGS measurements are crucial for accurate health assessments.
- Various dynamometers exist, necessitating validation for interchangeable use.
Purpose of the Study:
- To evaluate the inter-instrument agreement and criterion validity of the Baseline BIMS Digital Grip Dynamometer against the Jamar electronic dynamometer (Jamar+).
- To determine if HGS measurements from these two devices can be used interchangeably in clinical and research settings.
Main Methods:
- Seventy participants (aged 23-88) performed three HGS trials with each dynamometer in a randomized order.
- Intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated.
- New, manufacturer-calibrated dynamometers (Baseline BIMS and Jamar+) were utilized.
Main Results:
- An excellent overall ICC of 0.98 was observed between the two dynamometers.
- Minimal average differences were found for both the strongest attempt (0.68 kg) and the average of three attempts (0.37 kg).
- SEM and MDC values indicated low measurement error for the Baseline BIMS device compared to Jamar+.
Conclusions:
- The Baseline BIMS Digital Grip Dynamometer demonstrates high agreement and criterion validity when compared to the Jamar+.
- Results suggest that measurements obtained from both dynamometers can be used interchangeably.
- This facilitates consistent and reliable HGS assessments across different devices.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...