Thin Sectioning of Slice Preparations for Immunohistochemistry
Published on: April 28, 2007
Continuous specimen cooling during slicing and thickness measurement contributes to improved accuracy in image
Masanao Saio1, Raia Kushibiki2, Yuki Kanehira2
1Laboratory of Histopathology and Cytopathology, Department of Laboratory Sciences, Gunma University Graduate School of Health Sciences, Gunma, Japan. saio@gunma-u.ac.jp.
Abstract:
We measured section thickness (ST) after slicing using a film thickness meter and investigated the relationship between ST and the percent area of positive staining using computer-assisted image analysis.
Methods:
Sections were prepared from a paraffin-only block and formalin-fixed paraffin-embedded (FFPE) blocks containing fish sausage and human liver specimens. The ST was compared between the sections prepared with cooling using an ice pack (IP) or a continuous cooling device (CCD) paired with a sliding microtome set at an ST of 4 µm. The sections were stained with eosin or aniline blue, and the association between the percent area of positive staining and ST was determined using computer-aided analysis of images captured with a whole slide scanner.
Results:
The average STs of the paraffin-only block sections measured by four practitioners were 5.01-5.41 and 4.09-4.33 µm in samples prepared using an IP and a CCD, respectively. Therefore, subsequent analyses included sections prepared using the CCD. The ST of the tissue surface was significantly thinner than that of the paraffin surrounding the tissue section. Furthermore, the percent areas of positive staining for eosin and aniline blue were significantly correlated with ST in both the fish sausage and liver sections. The analysis of the ST and percent area of positive staining in 60 sections of the same block, which were categorized into quantiles based on ST, revealed a significant difference in the percent area of positive staining between the thicker and thinner sections.
Discussion:
Specimen sectioning should be performed with a CCD, ST should be measured before the staining of pathologic specimens prepared for quantitative analysis, and histologic examination should be performed using specimens with uniform ST.
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...