Gradual development of left bundle branch current of injury during left bundle branch pacing lead implantation
1Department of Cardiology, Kobari General Hospital, Chiba, Japan.
Abstract:
A larger left bundle branch (LBB) potential or LBB current of injury (COI) indicates a low LBB capture threshold in LBB pacing. During LBB pacing in an 85-year-old woman, achieving a low LBB capture threshold did not initially present with a larger LBB potential or LBB COI, but rather with a new initial negative deflection in a ventricular electrogram. LBB COI gradually developed over 7 min thereafter, which suggested that the lead tip had reached the left ventricular subendocardium. Therefore, this negative deflection may be the first sign to avoid further lead rotation.
Related Concept Videos
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrophysiology of Normal Cardiac Rhythm