A Multicenter MRI Protocol for the Evaluation and Quantification of Deep Vein Thrombosis
Published on: June 2, 2015
Thrombus discrimination with electron density: potential implications for non-contrast computed tomography imaging.
Gaston A Rodriguez-Granillo1, Juan Cirio2, Celina Morales3
1Department of Cardiovascular Imaging, Instituto Medico ENERI, Clinica La Sagrada Familia, Buenos Aires, Argentina.
Electron density (ED) imaging using dual-layer spectral computed tomography (CT) significantly improves thrombus visualization in vitro. This advanced imaging technique offers higher contrast-to-noise ratio for better identification of blood clots compared to conventional CT.
Area of Science:
- Medical Imaging
- Radiology
- Biomedical Engineering
Background:
- Dual-layer spectral computed tomography (CT) allows for direct estimation of electron density (ED) imaging.
- The clinical utility of ED imaging for thrombus identification remains underexplored.
- Accurate thrombus detection is crucial for effective treatment of acute ischemic stroke (AIS).
Purpose of the Study:
- To investigate the potential of ED imaging to improve thrombus identification compared to conventional imaging methods.
- To evaluate the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of ED imaging versus conventional and monoenergetic CT.
- To assess the feasibility of ED imaging for visualizing thrombectomy material in an in vitro setting.
Main Methods:
- Mechanical thrombectomy material from AIS patients was fixed and placed in polystyrene tubes.
- Samples were imaged using a dual-layer detector CT scanner, acquiring conventional CT (120 kVp), 40 keV monoenergetic, and ED images.
- Fifty-eight tubes (52 with thrombus, 6 controls) were analyzed for material presence, SNR, and CNR.
Main Results:
- ED imaging accurately identified thrombus material in all analyzed tubes, outperforming conventional CT which missed 3% of samples.
- ED imaging demonstrated a significantly higher signal-to-noise ratio (490.0±304.5) compared to conventional (10.4±7.0) and 40 keV (11.5±8.4) imaging.
- A substantial increase in contrast-to-noise ratio (CNR) was observed with ED imaging (37.8±29.1) versus conventional (4.3±4.3) and 40 keV (5.7±11.2) imaging (P<0.001).
Conclusions:
- In vitro, ED imaging significantly enhances the visualization of thrombus material compared to conventional and low monoenergetic CT imaging.
- The superior CNR provided by ED imaging suggests its potential for improved diagnostic accuracy in clinical applications.
- Further research is warranted to explore the clinical utility of ED imaging in diagnosing and managing conditions involving thrombus.
More Related Videos
13:10Combined Near-infrared Fluorescent Imaging and Micro-computed Tomography for Directly Visualizing Cerebral Thromboemboli
Published on: September 25, 2016
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...