Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Optimizing the Use of Dose Error Reduction Software on Intravenous Infusion Pumps
Kristen Hughes1, Madison Cole1, Deann Tims2
1Departments of Pharmacy and Pharmaceutical Sciences and.
Background And Objectives:
Intravenous infusions have the potential to cause significant harm in patients and are associated with a high rate of adverse drug events and medication errors. Infusion pumps with dose error reduction software (DERS) can be used to reduce errors by establishing safe infusion parameters. In 2019, a quality improvement project was initiated with the aim to increase DERS compliance from 46% to 75% at our specialty institution by October 1, 2022.
Methods:
An interdisciplinary group was tasked with improving compliance with DERS by identifying key drivers, including informed staff, engaged staff, and an accurate smart pump library. We used the Model for Improvement framework to guide this improvement project, and Plan-Do-Study-Act (PDSA) cycles were used to plan for interventions. PDSA cycles included drug library updates, education, and unit-level compliance reporting. Weekly average DERS compliance was monitored as the outcome measure, and weekly pump alerts per 100 infusions were monitored as a balancing measure; statistical process control charts were used to monitor measures from 2018 to 2022.
Results:
Over the course of 25 months, 8 PDSA cycles resulted in 5 centerline improvements from a baseline mean of 46% to a final mean of 78%. Pump alerts per 100 infusions decreased from 15.9 to 6.4 with the first PDSA cycle and then continued to decrease to 3.9 with subsequent interventions.
Conclusions:
Although features like DERS can help ensure safe medication administration, continuous improvement efforts to increase DERS compliance without increasing alert burden are needed to ensure that benefits of this technology are optimized.
More Related Videos
06:08Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Published on: March 11, 2017
08:51Multiple Intravenous Bolus Dosing and Invasive Hemodynamic Assessment in a Hypoxia-Induced Mouse Pulmonary Artery Hypertension Model
Published on: November 11, 2022
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Dosage Regimen: Fixed Dose
Fixed-dose regimens can be used for various routes of administration, including intravenous (IV) injections and oral medications. For IV administration, a predetermined amount of the drug is...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...