Choice of medium affects PBMC quantification, cell size, and downstream respiratory analysis
Ida Bager Christensen1, Lucas Ribas1, Maria Mosshammer2
1Xlab, Department of Biomedical Sciences, University of Copenhagen, Copenhagen, Denmark.
Mitochondrion
|May 8, 2024
Summary
Choosing the right medium is crucial for accurate high-resolution respirometry (HRR) of peripheral blood mononuclear cells (PBMCs). MiR05 medium may lead to underestimation of oxygen flux due to increased cell counts and aggregation.
Area of Science:
- Cellular bioenergetics
- Respirometry techniques
- Immunology
Background:
- High-resolution respirometry (HRR) is vital for assessing cellular energy metabolism.
- Standardized protocols for preparing peripheral blood mononuclear cells (PBMCs) for HRR are lacking.
- The choice of respirometry medium can significantly impact cell count, size, and bioenergetic measurements.
Purpose of the Study:
- To investigate the impact of four different media (MiR05, PBS, RPMI, Plasmax) on PBMC preparation and subsequent HRR.
- To determine the optimal medium for accurate assessment of intact PBMC bioenergetics using HRR.
- To provide data supporting the development of a standardized HRR protocol for PBMCs.
Main Methods:
- Intact PBMCs were prepared and analyzed in four different media: MiR05, PBS, RPMI, and Plasmax.
- PBMC count and size were determined for each medium.
- Oxygen (O2) flux was measured using high-resolution respirometry (HRR) with the Oxygraph-O2k instrument.
- Oxygen solubility factors for Plasmax and PBS were determined.
Main Results:
- Resuspending PBMCs in MiR05 resulted in a 21% higher cell count compared to PBS or Plasmax.
- MiR05 medium led to smaller cell size and increased cell aggregation.
- Higher cell counts in MiR05 caused an underestimation of O2 flux during HRR.
- Plasmax, PBS, and RPMI were found to be more suitable than MiR05 for HRR of intact PBMCs.
Conclusions:
- The choice of medium significantly affects PBMC count, size, aggregation, and subsequent HRR measurements.
- MiR05 is not recommended for HRR of intact PBMCs due to its impact on cell count and O2 flux.
- Plasmax, PBS, or RPMI are suggested as more appropriate media for standardized HRR protocols of intact PBMCs.
- Further optimization and standardization of HRR protocols for intact PBMCs are encouraged.