Large-scale Top-down Proteomics Using Capillary Zone Electrophoresis Tandem Mass Spectrometry
Published on: October 24, 2018
Top-Down Proteomics Analysis of Picogram-Level Complex Samples Using Spray-Capillary-Based Capillary
Zhitao Zhao1, Yanting Guo1, Trishika Chowdhury2
1Department of Chemistry and Biochemistry, University of Oklahoma, 101 Stephenson Parkway, Norman, Oklahoma 73019, United States.
We developed a highly sensitive spray-capillary capillary electrophoresis-mass spectrometry platform for top-down proteomics. This method enables ultrasensitive analysis of intact proteins in mass-limited biological samples, including single cells.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Molecular Biology
Background:
- Top-down proteomics is crucial for understanding biological systems in mass-limited samples like patient tissues and single cells.
- Existing top-down methods lack the sensitivity required for these challenging sample types.
- Capillary electrophoresis (CE) offers high resolution and low detection limits, making it suitable for mass-limited samples.
Purpose of the Study:
- To develop an ultrasensitive top-down proteomics platform for mass-limited biological samples.
- To enhance the sensitivity of electrospray ionization (ESI)-assisted spray-capillary devices for proteomics.
- To enable the analysis of intact proteins in complex samples down to the single-cell level.
Main Methods:
- Development of a spray-capillary-CE-MS platform incorporating a polyethylenimine (PEI)-coated capillary.
- Optimization of the spray-capillary inner diameter to improve sensitivity.
- Coupling nanodroplet-based sample preparation with the optimized CE-MS platform.
Main Results:
- Detection of over 200 proteoforms from 50 pg of E. coli lysate.
- Achieved high sensitivity, representing a significant advancement in top-down proteomics detection.
- Successfully identified 261 ± 65 and 174 ± 45 intact proteoforms from fewer than 50 HeLa and OVCAR-8 cells, respectively.
Conclusions:
- The modified spray-capillary CE-MS platform enables top-down proteomics analysis of picogram-level samples.
- This technology facilitates meaningful proteomic analysis of mass-limited samples, including single mammalian cells.
- Represents a breakthrough for ultrasensitive top-down proteomics in complex biological systems.
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...