Criterion (Concurrent) Validity and Clinical Utility of the Tongueometer Device
Brittany N Krekeler1,2,3, Anna Hopkins1, Meredith E Tabangin4
1Department of Otolaryngology-Head and Neck Surgery, University of Cincinnati College of Medicine, Cincinnati, OH.
Purpose:
Tongue manometry (i.e., tongue pressure measurement) is a commonly used assessment for patients with suspected oral-motor involvement in swallowing disorders. Availability of lingual manometry has changed in recent years, with the introduction of the Tongueometer device being a more affordable tongue manometry system. The purpose of this study was to test concurrent (criterion) validity of the Tongueometer compared to the current standard reference device, the Iowa Oral Performance Instrument (IOPI).
Method:
Adults without dysphagia were recruited for participation in this study. Standard lingual measurements (swallowing-related pressures, maximum isometric pressure [MIP], and maximum isometric endurance) were recorded, with the bulb anteriorly placed, with both devices, in a randomized order. The Bland-Altman method was used to determine concurrent (criterion) validity of these measurements compared to the clinical standard IOPI device. A recently available suggested corrective value by Curtis et al. (2023) was also applied, with comparisons made between devices both with and without the Curtis correction.
Results:
The final sample included 70 adult participants aged 20-89 years (Mage = 52.3 years). Measures with the Tongueometer device were significantly lower when compared with the same measures taken using the IOPI (p < .01) for all measures including MIP, endurance, and swallow pressures. The correction suggested by Curtis and colleagues did not ameliorate these differences.
Conclusions:
The Tongueometer lingual measurements were consistently lower compared to the IOPI. Clinical use of values taken with the Tongueometer device should be compared to normative data published for each specific device. Available features of each device (e.g., display, bulb texture, technology/application) should be considered when selecting which device to use with an individual patient.
Related Concept Videos
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Receiver Operating Characteristic Plot
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...