Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans
E Ø Buchhave1, C D Rasmussen1, R Lentz2
1Department of Diagnostic Radiology, Regionshospital Nordjylland, Frederikshavn, Barfredsvej 83, 9900 Frederikshavn, Denmark.
Introduction:
During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose (E), when positioning the arms below shoulder level in CT-TAP.
Methods:
An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate E. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.
Results:
Significant differences in E between all arm positions were seen (p = 0.005). The lowest E was obtained in position C, reducing E by 8.42%. Position A provided the best image quality but the highest E.
Conclusion:
This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest E. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.
Implication For Practice:
This study illustrates tendencies for objective image quality and E when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.
More Related Videos
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET