Related Experiment Video
Updated: Aug 15, 2026

High-resolution Functional Magnetic Resonance Imaging Methods for Human Midbrain
Published on: May 10, 2012
Improved MR imaging of the orbit at 1.5 T with surface coils
Abstract:
A method for obtaining localized high-resolution magnetic resonance (MR) images of the eye and orbit is demonstrated. The method uses modified surface receiver coils placed immediately adjacent to the anatomy to detect the MR signal. Surface coils provide enhanced sensitivity for imaging voxels close to the surface of the body while limiting the received patient-generated noise. The resulting improvement in signal-to-noise ratio allows for a reduction in the imaging voxel size to about 0.5 X 0.5 X 5 mm in scan times of 3.4-5 min. At this level of resolution, anatomic detail in the orbital region previously unobservable in MR images is seen.

