Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Efficacy of gonadal shielding in dose reduction for female pelvic radiography
Saki Nozoe1, Kei Terazaki1, Makoto Sasaki1
1Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
Abstract:
This study aimed to evaluate the dose reduction using gonad shielding (GS) during pelvic imaging. Three types of pelvic images (radiography, magnetic resonance and computed tomography) were fused to elucidate the three-dimensional relationship between the position of ovaries and GS. To estimate the dose received by the ovaries, the off-axis dose at any given depth was measured under two different imaging conditions using thermoluminescence dosemeters and a polymethyl methacrylate phantom. The mean ovarian depth was 8.4 cm. The mean estimated ovarian dose without an additional filter was 0.36 mGy without GS and 0.14 mGy with GS. The mean estimated ovarian dose with an additional filter was 0.24 mGy without GS and 0.10 mGy with GS. The efficacy of ovarian dose reduction should be evaluated based on the achieved ovarian dose, considering the ovarian depth and use of additional filtration, rather than the ovarian protection rate of GS.