Published on: July 18, 2014
Biventricular assist device implant using biatrial cannulation for restrictive cardiomyopathy
Fabian A Kari1, Jürgen Hörer1, Sebastian Michel1
1LMU University Hospital, Munich, Germany German Heart Center Munich, Munich, Germany European Children's Heart Center EKHZ, Munich, Germany.
Abstract:
Preoperative calculations showed that the 9-mm inlet, 6-mm outlet, 25-cc pump chambers and 65-73 bpm would be optimal for a 5-year-old patient suffering from restrictive cardiomyopathy, with a body surface area of 0.59 m2 (1.5 L/min flow for a cardiac index of 2.5). After re-sternotomy and standard bicaval cannulation for cardiopulmonary bypass, the procedure was performed under normothermic conditions and on the beating heart. Biventricular support was established with the Berlin Heart Excor using biatrial cannulation. For left atrial cannulation, induced ventricular fibrillation was used. The 9-mm inlet cannulas were inserted into the left and right atria, respectively. The 6-mm outlet cannulas were implanted using 8-mm interposition vascular grafts for the aorta and the main pulmonary artery, respectively. Cannulas were tunnelled through the epigastric space, with systems crossing outside of the body. The 25-cc chambers were used for both right ventricular assist device and left ventricular assist device support, which subsequently showed full emptying and filling.