Development and initial evaluation of a combustion-based sample introduction system for direct isotopic analysis of
Eduardo Bolea-Fernandez1, Ana Rua-Ibarz1, Jorge Alves Anjos1
1Ghent University, Department of Chemistry, Atomic & Mass Spectrometry - A&MS Research Group, Campus Sterre, Krijgslaan 281-S12, 9000, Ghent, Belgium.
Abstract:
High-precision isotopic analysis of mercury (Hg) using multi-collector ICP-mass spectrometry (MC-ICP-MS) is a powerful method for obtaining insight into the sources, pathways and sinks of this toxic metal. Modification of a commercially available mercury analyzer (Teledyne Leeman Labs, Hydra IIc - originally designed for quantification of Hg through sample combustion, collection of the Hg vapor on a gold amalgamator, subsequent controlled release of Hg and detection using cold vapor atomic absorption spectrometry CVAAS) enabled the system to be used for the direct high-precision Hg isotopic analysis of solid samples using MC-ICP-MS - i.e., without previous sample digestion and subsequent dilution. The changes made to the mercury analyzer did not compromise its (simultaneous) use for Hg quantification via CVAAS. The Hg vapor was mixed with a Tl-containing aerosol produced via pneumatic nebulization, creating wet plasma conditions, and enabling the use of Tl as an internal standard for correction of instrumental mass discrimination. Accurate and precise (0.10 ‰ 2SD, δ202Hg, n = 5) results were obtained for an in-house standard solution of Hg (20 ng Hg sample intake). Initial validation relied on the successful analysis of two solid certified reference materials of biological origin (BCR CRM 464 Tuna fish and NRC-CNRC TORT-3 Lobster hepatopancreas). It was shown that instrumental mass discrimination can be adequately corrected for by relying on the use of an aqueous Hg standard solution (NIST SRM 3133), without the need of matrix-matching. The novel setup developed thus allows for direct high-precision isotopic analysis of Hg in solid samples, thus enhancing the sample throughput. It is also suited for samples for which low amounts are available only and/or that are characterized by low Hg concentrations.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Atomic Emission Spectroscopy: Instrumentation