Excessive Heat Generation by Power-Driven Craniotomy Tools: A Possible Cause of Autologous Bone Flap Resorption
Satoru Shimizu1, Hiroyuki Koizumi2, Ichiyo Shibahara2
1Department of Neurosurgery, NHO Sagamihara National Hospital, Sagamihara, Kanagawa, Japan.
Background:
Bone flap resorption is an issue after autologous cranioplasty. Critical temperatures above 50°C generated by power-driven craniotomy tools may lead to thermal osteonecrosis, a possible factor in resorption. This ex vivo study examined whether the tools produced excessive heat resulting in bone flap resorption.
Methods:
Using swine scapulae maintained at body temperature, burr holes, straight and curved cuts, and wire-pass holes were made with power-driven craniotomy tools. Drilling was at the conventional feed rate (FR) plus irrigation (FR-I+), at a high FR plus irrigation (hFR-I+), and at high FR without irrigation (hFR-I-). The temperature in each trial was recorded by an infrared thermographic camera.
Results:
With FR-I+, the maximum temperature at the burr holes, the cuts, and the wire-pass holes was 69.0°C, 56.7°C, and 46.2°C, respectively. With hFR-I+, these temperatures were 53.1°C, 52.1°C, and 46.0°C, with hFR-I- they were 56.0°C, 66.5°C, and 50.0°C; hFR-I- burr hole- and cutting procedures resulted in the highest incidence of bone temperatures above 50°C followed by FR-I+, and hFR-I+. At the site of wire-pass holes, only hFR-I- drilling produced this temperature.
Conclusions:
Except during prolonged procedures in thick bones, most drilling with irrigation did not reach the critical temperature. Drilling without irrigation risked generating the critical temperature. Knowing those characteristics may be a help to perform craniotomy with less thermal bone damage.