Related Experiment Video
Updated: Jun 6, 2026

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Clinical Portrait of Cochlear Implantation in Patients With CHARGE Syndrome
Bokhyun Song1, Kyuhyeon Hwang1, Sung Min Koh1
1Department of Otorhinolaryngology-Head and Neck Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Objective:
To examine the clinical characteristics and auditory performance of patients with CHARGE syndrome following cochlear implantation (CI), as well as the prognostic factors affecting auditory outcomes.
Study Design:
Retrospective cohort.
Setting:
Tertiary academic center.
Methods:
A retrospective chart review was performed in patients with CHARGE syndrome who underwent CI from 2007 to 2022. The category of auditory performance (CAP) score was used to assess the CI outcomes, and factors that may affect the speech outcomes were also evaluated.
Results:
In 14 children with CHARGE syndrome, 22 CIs were performed, 6 unilaterally and 8 bilaterally. The mean age at CI was 25.9 months (range: 10-62). All patients had ear abnormalities and developmental delays, and cochlear nerve deficiency (CND) was present in all ears. At the last follow-up (mean: 49.6 months), the mean CAP score improved significantly compared to the preoperative measure (from 0.36 ± 0.81 to 3.21 ± 1.70, P = .001), with 6 patients (42.9%) achieving a CAP score of 4 points or higher. However, between the unilateral and bilateral CI groups, the final CAP score or change in CAP score was similar. Factors including age, coloboma, and CND did not significantly affect speech outcomes (all P > .05).
Conclusion:
Even though CHARGE syndrome features challenging anomalies, CI can be conducted safely and can offer effective contribution to significant speech improvement. Patients with CHARGE syndrome should be given the opportunity to undergo CI to maximize their audiological progress.