Comparison of a two-step Tempus600 hub solution single-tube vs. container-based, one-step pneumatic transport system
Marc Luginbühl1, Kathrin Frey1, Joanna Gawinecka1
1Institute for Clinical Chemistry, 229148 University Hospital and University Zurich , Zurich, Switzerland.
Objectives:
Efficient and timely transportation of clinical samples is pivotal to ensure accurate diagnoses and effective patient care. During the transportation process, preservation of sample integrity is crucial to avoid pre-analytical aberrations on laboratory results. Here, we present a comparative analysis between a two-step Tempus600 hub solution single-tube and a one-step, container-based pneumatic transport system (PTS) from Airco, for the in-house transportation of blood samples.
Methods:
Ten blood samples from healthy volunteers were split in 10 mL collection tubes filled at full or half capacity for transportation with the two PTS (about 250 m). To compare the impact of transportation, markers of hemolysis such as lactate dehydrogenase (LDH), potassium (K+), and the hemolysis index (HI), were determined. Additionally, differences in HI in routine samples and repeated transportation was investigated. To assess and compare the mechanistic impact profiles, we recorded the acceleration profiles of the two PTS using a shock data logger.
Results:
Transportation using the Tempus600 hub solution resulted in 49 and 46 % higher HI with samples filled to total or half capacity, respectively. Routine samples transported with the Tempus600 hub solution showed a higher median HI by 23 and 33 %. Additionally, shock logger analysis showed an elevated amount of shocks (6.5 fold) and shock intensities (1.8 fold).
Conclusions:
The Tempus600 hub solution caused an increased number of unreportable LDH or K+ results based on the hemolysis index. However, it was only statistically significant for LDH (p<0.01 and p<0.08) - while the comparisons for K+ were not statistically significant (p<0.28 and p<0.56).
Related Concept Videos
Principle of Linear Impulse and Momentum for a Single Particle: Problem Solving
Transport Number
Multimachine Stability
In analyzing the system, the nodal equations represent the relationship between bus voltages, machine voltages, and machine currents. The nodal equation is given by:
Multiple Pipe Systems
Series Configuration
In a series configuration, fluid flows sequentially from one pipe...
Problem Solving: Volume
Bioreactor Design and Operational System