Gas-phase ion mobility of protonated aldehydes in helium measured using a selected ion flow-drift tube
Maroua Omezzine Gnioua1,2, Patrik Španěl1, Anatolii Spesyvyi1
1J. Heyrovský Institute of Physical Chemistry of the Czech Academy of Sciences, Prague, Czechia.
Rationale:
In soft chemical ionization mass spectrometry, analyte ions are produced via ion-molecule reactions in the reactor. When an electric field E is imposed, the ion drift velocity vd determines the reaction time and the effective ion temperature. Agreement between experimental ion mobilities and theoretical predictions confirms the accuracy of the ion residence time measurement procedure.
Methods:
A selected ion flow-drift tube (SIFDT), an instrument with a chemical ionization source, was used to produce protonated aldehydes and selectively inject them into the resistive glass drift tube filled with He. Arrival-time distributions of ions were obtained using the Hadamard modulation. Reduced ion mobilities were then obtained at a pressure of 2 hPa in the E/N range of 5-15 Td. Theoretical ion mobility values were calculated using two methods: hard-sphere approximation and trajectory modelling.
Results:
The measured mobilities of three saturated and three unsaturated protonated aldehydes do not show substantial variation across the studied E/N range. Effective temperatures calculated using the Wannier formula from measured gas temperatures ranged from 300 to 315 K. Experimentally obtained values of the near-zero- E/N-reduced ion mobilities agree with both methods of calculations typically within ±3% standard deviation (maximum ±5%).
Conclusions:
The experimental SIFDT values of reduced mobilities in He of protonated aldehyde molecules generated from a chemical ionization source are in close agreement with two different theoretical methods based on the density functional theory calculations of ion geometries and partial atomic charges. Besides its fundamental importance, the ion mobility results validate the correct operation of the drift tube reactor and the ion residence time measurement procedure. Diffusion losses can also be determined from these results.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
16:40T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis
Published on: July 31, 2010
Related Concept Videos
Chemical Ionization (CI) Mass Spectrometry
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Gas Chromatography: Types of Detectors-II
Mass Analyzers: Common Types
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Capillary Electrophoresis: Instrumentation